Atrium OASIS
Dry Suction Water Seal Chest Drain
Set up
Step 1. Fill water seal to 2 cm line
Add 45 ml of sterile water or sterile saline via the blue
suction port located on top of the drain. For models
available with sterile fluid, twist top off bottle and insert
tip into suction port. Squeeze contents into water seal
until fluid reaches 2 cm fill line.
Step 2. Connect chest drain to patient(A)
Connect chest drain to patient prior to initiating suction.
Step 3. Connect chest drain to suction
Attach suction line to the blue suction port on top of
chest drain.
Step 4. Turn suction source on
Increase suction source vacuum to –80 mmHg or
higher. Suction regulator is preset to –20 cmH
2
O.
Adjust as required.
Have a question or need help in a hurry?
Call Maquet toll free at 1-800-528-7486.
What to check during system operation
Verifying suction operation via the suction
monitor bellows
The bellows located in the
suction monitor window will
expand only when suction
is operating. The bellows
will not expand when
suction is not operating
or is disconnected. The s
mark allows quick and easy
confirmation of vacuum
operation over a wide
range of continuously
adjustable suction control
settings.
Increase vacuum source when bellows is not
expanded to smark
If the bellows is observed to be expanded, but less
than the smark, the vacuum source pressure must be
increased to –80 mmHg or higher.
Changing suction pressures
Changing suction pressure
is accomplished by adju-
sting the rotary dry suction
control dial located on the
side of the drain. Dial down
to lower the suction pres-
sure and dial up to increase
the suction pressure.
NOTE: When changing suction pressure from a higher
to lower level, use of the manual high negativity vent
after regulator adjustment will reduce excess vacuum
pressure down to the lower prescribed level.
Verifying water seal operation
The water seal must be filled and maintained at the
2 cm level to ensure proper operation and should be
checked regularly when used for extended periods. As
required, additional water may be added by a 20 gauge
or smaller needle and syringe via the grommet located
on the back. Fill to the 2 cm line.
Placement of unit
Always place chest drain below the patient’s chest in an
upright position. To avoid accidental knock-over, open
the floor stand for secure placement on floor or hang the
system bedside with the hangers provided.
Recording drainage volume
The collection chamber incorporates a writing surface
with easy-to-read fluid level graduations. Please refer
to individual product inserts for specific model gradu-
ations.
Observing water seal for
patient air leaks
The chest drain offers air leak
detection with rapid air leak
assessment and enhanced
visibility due to the tinted
water. When air bubbles are
observed going from right to left in the air leak monitor,
this will confirm a patient air leak.
Continuous bubbling in the bottom of the water seal
air leak monitor will confirm a persistent air leak.
Intermittent bubbling in the air leak monitor with float
ball oscillation will confirm the presence of an intermit-
tent air leak.
No bubbling with minimal float ball oscillation at bot-
tom of the water seal will indicate no air leak is present.
Observing graduated water seal column for
changes in patient pressure
Patient pressure can be
determined by observing the level
of the blue water and small float
ball in the graduated water seal
column. With suction operating,
patient pressure will equal the
suction control setting plus the
graduated water seal column level.
For gravity drainage (no suction) patient pressure will
equal the graduated water seal column level only.
Graduated air leak monitor
For those models with a graduated air leak monitor,
air leak bubbling can range from 1 (low) to 5 (high). Air
bubbles create an easy to follow air leak pattern for
monitoring patient air leak trends.
High negativity float valve
The high negativity float valve,
with its controlled release action,
enables the thoracic patient to
draw as much intra thoracic pres-
sure as is required during each
respiratory cycle. During prolon-
ged episodes of extreme negative
pressure, a controlled release
system will automatically relieve excess vacuum to the
lower, prescribed pressure level.
Manual high negativity vent
To lower the height of the water seal column or to
lower patient pressure when connected to suction,
depress the manual
vent located on top of
the drain until the float
valve releases and the
water column returns to
the desired level. Do not
use manual vent to lower water seal column when
suction is not operating or when the patient is on
gravity drainage.
Positive pressure protection
The positive pressure
release valve, located on top
of drain, opens to release
accu mulated positive
pressure. Do not obstruct
the positive pressure
release valve.
G
I
E
L
K
J
A
B
D
C
P
O
N
M
Bellows must be expanded to s
mark or beyond for a –20 cmH
2
O
or higher regulator setting.
Do not use when suction is not operating.
A Dry Suction Regulator
B Water Seal Chamber
C Air Leak Monitor
D Collection Chamber
E Suction Monitor Bellows
F Easy-to-Grip Handle
G Positive Pressure Release Valve
H Suction Port
I Manual High Negativity Vent
J Multi-position Hangers
K In-line Connector
L Needleless Access Port
M Patient Pressure Float Ball
N Swing Out Floor Stand
O Patient Tube Clamp
P Patient Connector
F
H
Troubleshooting
Sampling patient drainage
Sampling of patient drainage must be in accordance
with approved hospital infection control standards.
Selected models include a needleless Luer port on the
patient tube connector for sampling patient drainage.
Alcohol swab the Luer port prior to syringe attach-
ment (no needle). Fluid samples can also be taken
directly from the patient tube by forming a temporary
dependent loop and inserting a 19 gauge needle at an
oblique angle. Alcohol swab the patient tube prior to
inserting syringe at a shallow angle. Do not puncture
patient tube with an 18 gauge or larger needle.
System disconnection
For models equipped with an in-line connector, close
the patient tube clamp prior to disconnecting the
chest drain patient tube from patient. Clamp off all
indwelling thoracic catheters prior to disconnecting
chest drain from patient.
System disposal
Disposal of chest drain and its contents should be in
accordance with all applicable regulations.
Have a question or need help in a hurry?
Call Maquet toll free at 1-800-528-7486.
M Atrium Ocean Chest Drain is manufactured by Atrium Medical Corporation, 5 Wentworth Drive, Hudson NH 03051 603-880-1433. • Protected by the following international and U.S. patent(s): http://patents.maquet.com. • Y CAUTION: Federal (US) law restricts
this device to sale by or on the order of a physician. • Maquet is a Trademark and/or Registered Trademark of Maquet GMBH, its subsidiaries, and/or affiliates in the United States and/or other countries. • Maquet is registered in U.S. Patent and Trademark Office.
• Ocean and Atrium are Trademarks and/or Registered Trademarks of Atrium Medical Corp., its subsidiaries, and/or affiliates in the United States and/or other countries. • Ocean and Atrium are registered in U.S. Patent and Trademark Office. All other company or
product names are the trademarks or registered trademarks of their respective holders. • Copyright 2015 Atrium Medical Corp. or its affiliates. • All rights not expressly granted are reserved. • 03/15 • Part No. 010394 Rev AA Letter. Refer to Instructions for Use for
current indications, warnings, contraindications, and precautions.
What to check during system operation
How do I determine patient pressure with a dry suc-
tion chest drain?
Whether using a traditional wet or dry suction operating
system, one cannot overemphasize the importance
of the graduated water seal column when it comes to
diagnosing the patient’s condition or monitoring normal
system operation. Patient pressure can be determined
by observing the level of the blue water and small float
ball in the graduated water seal column. With suction
operating and the bellows expanded across the suction
monitor window, patient pressure will equal the suction
control setting (read directly from the regulator dial) plus
the graduated water seal column level. For example,
when the suction monitor bellows is expanded to the
s mark or beyond to confirm a –20 cmH
2
O suction
setting, and the graduated water seal column reads
–15 cmH
2
O, patient pressure is –35 cmH
2
O (–20 cmH
2
O
+ –15 cmH
2
O = –35 cmH
2
O). For gravity drainage (no
suction) patient pressure will equal the graduated water
seal column only.
What should I do when the suction monitor bellows
is not expanded to the s mark when the regulator is
set at -20 cmH
2
O or higher?
The position of the bellows across the suction monitor
window will alert the operator that the suction source
has fallen below the minimum operating range for the
prescribed suction control setting. Simply increase the
vacuum source to –80 mmHg or higher. The suction
monitor bellows must expand to the smark or beyond
for –20 cmH
2
O or higher suction regulator setting.
What should I do when the bellows does not fully
expand to the smark after I increase the suction
source vacuum?
Dry suction chest drains require higher levels of vacuum
pressure and air flow from the suction source to operate
properly at each suction control setting as compared
to traditional water controlled operating systems. The
suction source should provide a minimum vacuum
pressure of –80 mmHg at 20 liters of air flow per minute
for chest drain operating efficiency at a suction control
setting of –20 cmH
2
O. The suction source should be
greater than –80 mmHg when multiple chest drains are
connected to a single suction source. If the bellows
does not fully expand to the smark, it may simply
be that the suction source is not functioning to its full
potential to provide the minimum vacuum pressure or
air flow required to “drive” the suction control regulator.
Additionally, conditions may exist that can reduce, or
“restrict” air flow from the suction source. A restrictive
clamp, connector, or kink in the suction line tubing can
potentially “starve” the chest drain of air flow. A leak in a
connection or wall canister, along with extensive lengths
of suction tubing can also reduce air flow to the unit.
To troubleshoot this situation, first check to be sure
that all connections are air-tight. Inspect the suction
tubing and connections for possible cracks, leaks,
kinks, or occlusion. You may need to simply bypass a
“leaky” wall canister. Try connecting the chest drain to
a different suction source or wall regulator. When mul-
tiple chest drains are “Y” connected to a single suction
source, if possible, reconnect the drains to separate
suction sources. Finally, replace the chest drain if you
suspect the unit is cracked or damaged.
Does the bellows need to expand beyond the s
mark for a –10 cmH
2
O regulator setting?
No. For a regulator setting less than –20 cmH
2
O suction
(–10 cmH
2
O), any observed bellows expansion across
the monitor window will confirm suction operation. The
bellows need not be expanded to the smark for suc-
tion pressures less than –20 cmH
2
O, just visibly expan-
ded to confirm suction operation.
How do I confirm my patient has an air leak when:
There is no bubbling in the water seal?
If there are no air bubbles observed going from right to
left in the air leak monitor, there is no patient air leak.
In order to confirm that your patient’s chest catheter
is patent, temporarily turn suction off and check for
oscillation of the patient pressure float ball in the water
seal column coinciding with patient respiration.
Bubbling is present in the water seal?
Whenever constant or intermittent bubbling is present
in the water seal air leak monitor, this will confirm an
air leak is present. Oscillation of the patient pressure
float ball at the bottom of the water seal without
bubbling will indicate no apparent air leak. Bubbling
from right to left must be present to confirm an air
leak. To determine the source of the air leak (patient or
catheter connection), momentarily clamp the patient
tube close to the chest drain and observe the water
seal. If bubbling stops, the air leak may be from the
catheter connections or the patient’s chest. Check
the cathe ter connectors and patient dressing for a
partially withdrawn catheter. If bubbling continues after
temporarily clamping the patient tube, this will indicate a
system air leak requiring system replacement.
What does it mean when the small float ball is
located at the bottom of the water seal column?
If the small float ball is located and oscillating at the
bottom of the water seal column with no bubbling,
there is no apparent patient air leak. However, the water
seal should be carefully monitored for the presence of
an occasional or intermittent air leak.
Is it normal for the patient pressure float ball to
fluctuate up and down (tidal) near the bottom of the
water seal column?
Yes. Once your
patient’s air leak is
resolved, you will
generally observe
moderate tidaling
in the water seal
column. Increases in
intrathoracic pressure
will cause the water
level to rise (the ball
rises) during patient
inspiration and will lower or decrease (the ball drops)
during expiration. This diagnostic tool will help to
confirm patency of your patient’s catheter(s).What
happens when the water rises to the top of the
water seal float valve?
The water seal column is a diagnostic manometer for
monitoring your patient’s intrathoracic pressure. When
intrathoracic pressures increase, causing the water
to rise to the top of the water seal float valve, the ball
floats up and “seats” up against a valve seat. This
valve seat has been engineered to allow a specific
amount of water to pass through it during a defined
amount of time. When vacuum pressures greater
than –20 cmH
2
O on gravity or –40 cmH
2
O on suction
occur for an extended period of time, water will pass
through the valve seat and float valve to allow the
water seal to release automatically. The benefit of the
controlled release design is that during normal or deep
inspiration, the float valve will float up and down with
each respiratory cycle, not allowing the water seal to
release. This enables thoracic patients to draw as much
intrathoracic pressure as they may require during each
respiratory cycle.
How do I lower the water seal column?
Changes in your patient’s intrathoracic pressure will
be reflected by the height of the water in the water seal
column. These changes are usually due to mechanical
means such as milking or stripping patient drainage
tubes, or simply by deep inspiration by your patient after
all air leaks have subsided. If desired, the height of the
water column and patient pressure can be reduced by
temporarily depressing the filtered manual vent located on
top of the drain, until the float valve releases and the water
column lowers to the desired level. Do not lower water
seal column when suction is not operating, or when
patient is on gravity drainage.
If the chest drainage system has been knocked
over, can I use it and what should I do?
After a chest drainage system has been knocked over,
set it upright and immediately check the fluid level of the
water seal for proper volume. The drain provides a con-
venient diaphragm for access by a 20 gauge or smaller
needle and syringe to adjust the water level in the water
seal chamber, if required. Alcohol swab the needle access
area and aspirate any overfill that may have occurred.
If the water seal has an inadequate fluid level, simply
replace the lost volume. If a significant amount of
blood has entered the water seal, it may be advisable
to change the system for a new one.
Not enough vacu um for
-20 cmH
2
O or higher suction
control setting.
Normal suction operation for
- 20 cmH
2
O or higher.
Increase suction source
to -80 mmHg or higher.
Maquet Medical Systems USA | 45 Barbour Pond Drive, Wayne, NJ 07470 | www.maquetusa.com MCV00033028 REVA